JMNH LLC
Contents

General educational reading

A reading changes over time.

A changing indication may describe the thing being measured. It may also reflect a change in the instrument itself. Time makes that distinction important.

01A changing instrument

Instrumental drift is a change in indication over time caused by changes in an instrument’s metrological properties. It is distinct from a real change in the measured quantity or a change in a recognized influence quantity, such as ambient temperature.

The distinction matters because several effects can produce similar-looking records. A gradual slope is an observation, not a diagnosis. Instrument history, environmental records and suitable reference checks can help investigate which explanation fits.

02Keep a reference in view

Repeated checks against a suitable stable reference can reveal changes in the measurement system. Their interpretation still depends on the reference, the check procedure and the uncertainty of the comparison. A check is evidence within those conditions.

A useful record preserves when a check occurred and whether anything was adjusted. Correcting later readings without retaining that history can make a sequence difficult to interpret. Calibration intervals and acceptance criteria should follow the application and relevant technical guidance rather than a universal calendar rule.

What independent observation would help separate a changing object from a changing instrument?

Further reading: JCGM vocabulary: instrumental drift.