01A documented connection
Metrological traceability belongs to a measurement result. It relates that result to a reference through an unbroken, documented chain of calibrations, each contributing uncertainty. An instrument label by itself does not describe the complete chain.
The reference may be a realization of a measurement unit or another clearly specified reference. Documentation needs to make the connection understandable. Knowing where an instrument was purchased is a different kind of history and does not establish metrological traceability.
02Connection is not suitability
A traceable result may still have uncertainty too large for a particular decision. Traceability and fitness for purpose therefore need separate consideration. The measurement procedure and conditions also matter when applying calibration information to a result.
When reading a report, look for the stated reference, the associated uncertainty and the supporting calibration information. A broad endorsement or a familiar institutional name cannot replace those details. The useful question is how this particular result is connected, and what that connection supports.
Can the reference connection be followed for the result in front of you?
Further reading: NIST: metrological traceability questions.